朝夕说 · 英语阅读

Measurement Equivalence of On-Scalp OPM-MEG and Cryogenic MEG for Auditory and Somatosensory Cortical Mapping Across Development

C2科学208 词约 1 分钟

Wearable optically pumped magnetometer magnetoencephalography (OPM-MEG) reduces sensor-to-cortex distance compared with conventional cryogenic SQUID-MEG, but whether the two technologies yield equivalent neurophysiological conclusions remains unclear. We recorded auditory and somatosensory evoked fields in 18 participants (10-45 years) using a 128-sensor FieldLine HEDscan OPM-MEG system and a 275-channel CTF SQUID-MEG system within the same individuals. Equivalent current dipole source models were estimated using identical preprocessing and modeling procedures and compared using paired permutation testing.

Both systems localized canonical auditory and somatosensory cortical generators with matched peak latencies and modest cross-system spatial differences. Auditory sources showed a consistent medial bias in SQUID-MEG localization, whereas somatosensory sources exhibited a small systematic offset ([~]4 mm), indicating stable coordinate differences rather than localization error. Dipole moments were larger for SQUID-MEG and goodness-of-fit higher for OPM-MEG; however, the increased moment was explained by a medial localization bias, demonstrating inverse-model effects rather than physiological disagreement. Auditory dipole moment increased with age in both systems, whereas somatosensory responses showed no age-related change. Together, these observations indicate preserved developmental physiology across platforms.

Gaetz, W. et al. · CC-BY 4.0

朝夕说 · 听说读写背单词 · 赣ICP备2026010754号